Description
Universal tension and compression load cell with high accuracy.
Features- Universal tension and compression load cell with high accuracy.
- Wide range capacity: 20 kN (2000 kg) to 200 kN (20 tons)
- Tension and compression
- Easy to install
- "S" Compact design
- Protection IP 67
- Material: nickel plated steel
- Available options (non exhaustive list):
- High service temperature (up to 180° C)
- ATEX intrinsic safety Ex II 1GD Ex ia IIC T6 or T4 Ga Ex ia IIIC T80° C Da
- Cable output by cable gland
- Also available as standard reference load cells following ISO 376
Specifications
Specifications
Accuracy
- 0.03 % of Full Scale
- 0.05 % of Full Scale
- 0.1 % of Full Scale
Applications
- Conveyors
- Hoppers
- Measurement
- Reactors
Approvals & Certifications
- ATEX Intrinsic Safety Ex II 1 GD Ex ia IIC T6 or T4 Ga Ex ia IIIC T80° C Da
- CE
Cable Length
- 10 to 32 feet (3 to 10 metres)
Capacity
- 2000 kg to 20 Tons
Electrical Connectors
- Cable
Enclosure, Body Material
- Nickel-Plated Steel
Environmental Protection
- IP67
Excitation Voltage
- 10 Vdc
Linearity
- Linearity Error: < ± 0.025 % of Full Scale (Class 0.03) to < ± 0.1 % of Full Scale (Class 0.1)
Load Limit
- 120 % of Full Scale
Operating Temperature
- 180° C (356° F)
Repeatability
- Error: < ± 0.015 % (Class 0.03) to < ± 0.03 % of Full Scale (Class 0.1)
Resistance
- Input/Output: 352 ± 2 ohms
Sensitivity
- Tolerance: < ± 0.2 % (Class 0.03) to < ± 0.3 % (Class 0.1)
Service Temperature
- -30° to 70° C (-22° to 158° F)
Static Lateral Force
- 50 % of Full Scale
Storage Temperature
- -50° to 85° C (-58° to 185° F)
Temperature Coefficient
- < ± 0.015 % to < ± 0.05 % of Full Scale/10° C (18° F)
- < ± 0.023 % to < ± 0.035 % of Full Scale/10° C (18° F)
Zero Balance
- ± 0.02 mV/V