Description
The Multi-Function Calibrator CEP6000 provides a feature set unmatched in high accuracy, hand-held calibrators in its price range. The CEP6000 provides the functions and accuracy associated with fixed installation, laboratory instruments, and has everything needed for virtually any calibration task. Measure and source thermocouples, resistance thermometers, current, voltage, frequency as well as pressure measurement and source pulse trains is supported by the CEP6000. External pressure modules can be connected to an interface. The built-in isolated mA/V-measuring channel allows a complete transmitter calibration.
Features- Highest accuracy in class of up to ±0.015 % of reading.
- Measuring and simulation of thermocouples (13), resistance thermometers (13), resistance, current, voltage, frequency, pressure and pulse trains.
- Isolated mA/V measuring channel enables complete transmitter calibration (measuring and simulation at the same time).
- Entry of customer-specific resistance thermometer coefficients.
- Robust design and simple operation.
Specifications
Specifications
Display
- 2-part, each with 10 Digits and 8 mm character size
Input
- 6 banana plug Inputs for Electrical Parameters, resistance thermometers and thermocouples
Resistance Thermometer (RTD)
- Pt100
- (385, 3926, 3916)
- Pt200
- Pt500
- Pt1000
- Ni120
- Cu10
- Cu50
- Cu100
- Pt10
- Pt50
- YSI400
Thermocouple Types
- XK
- BP
- C
- N
- U
- L
- J
- K
- T
- E
- R
- S
- B
Voltage Signal
- Input: DC 30 V
- output: DC 20 V
Current Signal
- Input: DC 24 mA
- Output: DC 24 mA
Resistance
- 0 to 4,000 Ω
Frequency / Pulse
- 2 CPM ... 10 kHz
Pressure
- dependent on Pressure module
Voltage Supply
- DC 24 V
Resistance Thermometer Frequency Response
- 5 ms, works with all pulsed transmitters
Customer-Specific Resistance Thermometers
- Entry of customer-specific resistance thermometer coefficients
Functions
- automatic step function
Interface
- RS-232, USB with optional Serial Adapter
Power Supply
- 4 x 1.5 V AA batteries
Battery Life
- 20 hours
Battery Status Indication
- Icon in display for Low battery level
Operating Temperature
- -10° to +50° C
Storage Temperature
- -20° to +70° C
Relative Humidity
- 0 to 90 % Relative Humidity, Non-Condensing
Temperature Coefficient
- 0.003 % Full Scale/°C, outside of 23° C ± 5° C
Applications
Applications
- Calibration service companies and service industry
- Instrument and control workshops
- Industry (laboratory, workshop and production)
- Quality assurance