Model DPGT43.100, DPGT43.160 Differential Pressure Gauge with Output Signal
Description
Wherever the process pressure has to be indicated locally and, at the same time, a signal transmission to the central control or remote centre is desired, the model DPGT43 intelliGAUGE® (patent, property right: e.g. DE 202007019025) can be used.
Features- No configuration necessary due to “plug-and-play”.
- Signal transmission per NAMUR.
- Differential pressure measuring ranges from 0 … 16 mbar.
- Easy-to-read Analog display with nominal sizes 100 and 160.
- Individual, non-linear characteristic curves (e. g. x or √x for flow measurement).
Specifications
Specifications
Design
- ASME B40.100 & EN 837-3
Sizes
- 4” or 6” (100 or 160 mm)
Accuracy Class
- ± 2/1/2 % of Span (ASME B40.100 Grade A)
Range
- Scale Range 0/6.5”H₂O: full scale length approximately 180° or other equivalent units of pressure or vacuum
- 0/6.5 “H₂O up to 0/400 psi
Overpressure Safety
- See table on page 4
Operating Temperature
- Medium: +212° F (+100° C) Max.
- Ambient: -4° F to +140° F (-20° C to +60° C)
Temperature Error
- Additional error when temperature changes from reference temperature of 68° F (20° C) ± 0.5 % for every 18° F (10° C) ris ing or falling. Percentage of Span.
Measuring Chamber with Pressure Connection (Wetted)
- 2 x 1/4” NPT Female
- Lower Mount (LM)
- Material: 316L Stainless Steel
Pressure Element (Wetted)
- 100”H₂O: NiCrCo-alloy (Duratherm)
Venting of the Media Chambers (Wetted)
- 316L Stainless Steel for Ranges 160 “H₂O)
Sealing Bellow (Wetted)
- 316L Stainless Steel
Movement
- Copper Alloy
Dial
- White Aluminum with Black Lettering
Pointer
- Black Aluminum, Adjustable
Case
- Stainless Steel, solid-front, blow-out back
Window
- Laminated Safety Glass
Cover Ring
- Bayonet Ring, Stainless Steel
Weather Protection
- NEMA 4X / IP54 per EN 60 529 / IEC 529 (with Liquid filling NEMA 6 / IP65)
Mounting
- (+) High Pressure, (-) Low Pressure
- According to engraved symbols on the body
Mounting Options
- Drilled Mounting holes in the measuring flange
- Rigid tailpipes
- Panel Mounting flange (Optional)
- Pipe or surface Mounting bracket (Optional)
Applications
Applications
- Acquisition and display of processes
- Output signals 4 ... 20 mA, 0 ... 20 mA, 0 ... 10 V for the transmission of process values to the control room
- For measuring points with increased differential overpressure
- Easy-to-read, analogue on-site display needing no external power
- Safety-related applications