3115-ISO High-Capacity Compression Standard Reference Force Transducer
Description
Standard reference force transducer specially designed according to the norm ISO 376(1) (Class "1", "05" and "00").
Features- Broad range of capacities (up to 30 MN)
- Complete range of load accessories
- Protection: IP 65
- Material: nickel plated alloy steel
- Options:
- Digital output RS-232C, RS-485 or USB
- Custom made dimensions
- ASTM E74 compliant
Specifications
Specifications
Accuracy
- Class 00
- Class 05
- Class 1
Applications
- Control of Testing Machines According to ISO 7500
Approvals & Certifications
- ISO 7500
Approvals & Certifications
- CE
Cable Length
- 19.96 to 39.37 feet (6 to 12 metres)
Capacity
- 30 to 30,000 KN
Communication
- RS-232
- RS-485
- USB
Electrical Connectors
- DIN 45322 Female-Male Connector
Enclosure, Body Material
- Nickel-Plated Alloy Steel
Environmental Protection
- IP65
Excitation Voltage
- 10 Vdc - Reference
- 3-12 Vdc - Nominal Range
Insulation Resistance
- > 5,000 megaohms at 50 V
Load
- > 300 % of Full Scale - Breaking Load
Load Limit
- 110 % of Full Scale - Safe Load Limit
Repeatability
- Relative Error with Rotation: < ± 0.2 % R.O. (Class 1) to < ± 0.05 % R.O. (Class 00)
- Relative Error without Rotation: < ± 0.1 % R.O. (Class 1) to < ± 0.025 % R.O. (Class 00)
Resistance
- 350 to 702 ± 2 ohms - Input & Output Resistance
Sensitivity
- 1.5 mV/V - Nominal Sensitivity
Service Temperature
- -30° to 70° C (-22° to 158° F)
Storage Temperature
- -50° to 85° C (-58° to 185° F)
Temperature Coefficient
- Sensitivity: < ± 0.035 % of Full Scale/10° C (Class 1, & Class 05) to < ± 0.015 % of Full Scale/10° C (Class 00)
- Zero Signal: < ± 0.03 % of Full Scale/10° C (Class 1, & Class 05) to < ± 0.023 % of Full Scale/10° C (Class 00)
Temperature Range
- -10° to 45° C (14° to 113° F) - Nominal Temperature Range
- 20° C (68° F) - Reference Temperature
Zero Balance
- ± 0.02 mV/V
Zero Offset
- Zero Shift after Loading: < ± 0.05 % of Full Scale (Class 1) to < ± 0.012 % of Full Scale (Class 00)